B8199999EU Ion Source Filament Edwards ELD500 Wet / Dry
Part identification
Description
B8199999EU Ion Source Filament — Edwards ELD500 WET / DRY / FLEX Leak Detectors Replacement ion source filament for the Edwards ELD500 in all its variants — WET, DRY and FLEX. The cell carries two filaments, one in service and one on standby, and the controller switches automatically when the first fails. That is a useful safety net and also a trap: a machine running on its standby filament looks perfectly healthy until the second one goes, and then it stops mid-shift. Check which filament your ELD500 is currently running on. If it has already switched, order two — you are one failure away from an unplanned stop. The same part number, B8199999EU, is used on the Pfeiffer HLT series (HLT260 / HLT550 / HLT560). One filament, two brands. Supplied as a Taixing-manufactured equivalent. Fit with clean gloves and recalibrate against a calibrated leak afterwards.
Technical data
| Cross-reference part no. | B8199999EU |
|---|---|
| Fits leak detectors | Edwards ELD500 · ELD500 WET · ELD500 DRY · ELD500 FLEX (dual filament) |
| Type | Ion source filament |
| Supplied by | Taixing |
Need a confirmed configuration?
Share your application, target vacuum and voltage. We will match the pump or part and return a formal quote.